CS328A-XS
Fully configured USB 2.0 oscilloscope - 14 bit 100 MSa/s, 2 analog channels, 8 digital inputs, External Trigger, 8M samples storage for each of the 2 analog and 8 digital channels, CS700 Signal Generator (0-10MHz, Sine, Square, Triangle). Complete with the CS300 application software, 2 x CS1001 and CS1004/5 probe sets. The front panel ground is continuous via the USB ground to the PC.
Cleverscope Model CS328A is a USB connected, PC hosted oscilloscope and spectrum analyser with PC based application software.
CS328A-XS Acquisition Unit
-
Two 14 bit analog channels sampling simultaneously at 100 MSa/sec. AC or DC coupled.
-
Streaming transfer to PC disk at up to 1.5MSPS with 14 bit ENOB, for up to 200G Samples.
-
USB 2.0 Interface to the PC or 10/100Mbit ethernet, depending on option
-
Gain automatically set from 20 mV FSD to +/-200V FSD by choosing graph view and probe switch setting.
-
Offset automatically set to correspond to the selected amplitude axis view. As an example a graph with amplitude axis of 2.40 to 2.42V automatically chooses a 20mV range offset by +2.40V - good for looking at small signals in single supply op-amp applications.
-
Analog triggering of the waveform in view with a resolution of 1% of the display height. The analog trigger may optionally be conditioned with a low pass, high pass or noise filter.
-
One external trigger, threshold adjustable from 0 to ±18V in 40 mV increments.
-
Eight digital inputs sampling at 100 MSa/sec, threshold adjustable from 0 to 10 V in 10 mV increments.
-
A dual hardware trigger system based on a rising or falling edge on any input signal, optionally qualified by a user defined digital input combination, triggering when one or both trigger conditions are met, and the time between triggers is less than, in or more than a user defined range, or the trigger count equals a user defined value.
-
A rear panel I/O connector with a 100 Mbit/sec bi-directional LVDS/RS422 link, and three RS422 outputs defaulting to sampling started, trigger received and sampling stopped.
-
Each channel (two analog, ext trigger and eight digital) includes 8M samples of storage, providing up to 40 ms of simultaneous storage for each channel, with 10 ns resolution.
-
The sample storage may be allocated as between 2 to 3000 frames varying in size from 4M to 2000 samples. These may be used as a history store for reviewing previously captured signals, or to capture up to 3000 trigger events with a minimal inter-frame delay, while maintaining time relative to the first trigger for all succeeding frames.
-
20 MHz 5th order Anti-alias filter for improved Spectrum Analysis performance.
-
Triggered LED and Power LED on the front panel.
-
Input power range from 6 ~ 12V, 6W provided by a universal mains adaptor.
-
Low jitter (1 ps rms) sampling clock for 70 dB spurious free dynamic range.
-
Self calibration to ensure DC performance specifications are met.
-
Enclosure size: 153 x 195 x 35 mm.
CS300 Application Software
-
Separate, freely moveable and resizable windows to display the signal, a zoomed signal view, and the frequency spectrum of the signal, and control panel.
-
The zoomed signal view optionally tracks the signal view cursor.
-
Spectrum analysis with a variety of conditioning windows and display in log or linear format.
-
Each signal window includes a time/amplitude tracer, and two markers for comparison purposes. Colours are user definable.
-
Signal averaging (exponential, block and peak hold) and low pass filtering.
-
Signal measurement, including Peak to Peak, RMS, DC, pulse width, period and frequency.
-
Copy and Paste graphic or data to other applications.
-
Save and Open from disk.
-
User defined units, signal names and scaling (offset and gain).
-
Text annotation of each graph.
-
Maths equation builder with realtime Matlab transfer.
-
Protocol Decoding for SPI, UART, I2C and bus.
CS328A-XS Equipment List
The following equipment is included with the CS328A-XS:
-
1 x CS328A-XS Oscilloscope Acquisition Unit with 14 bit, 8 MSample capture
-
1 x CS700A 0-10MHz Signal Generator factory fitted
-
1 x Power Supply Unit and lead
-
2 x CS1001 1x/10x 250MHz probe kit
-
1 x CS1004/CS1005 8 bit 100MHz digital input pods including probes
-
1 x USB cable
-
1 x CD that includes CS300 Application Software, CS328A User Manual, sample files, videos and white papers